Note:
a) The following accuracy specifications are valid at 23 degrees C,
+/- 5 degrees C, Relative Humidity less than 75%
b) The specifications are in the form " +/- (x % of reading + dgt)" where dgt is
"Least Significant DIGIT".
DC Voltage
Range Resolution Accuracy
200mV 0.1mV
2000mV 1mV
20V 10mV
200V 100mV
± (0.5% rdg + 2 digits)
500V 1V ± (0.8% rdg + 2 digits)
Input Impedance: All ranges are 10M
Overload Protection: 500Vrms
AC Voltage
Range Resolution Accuracy
200V 100mV
500V 1V
± (1.2% rdg + 10 digits)
Input Impedance: All ranges are approx. 5M
Frequency: 40Hz to 400Hz
Overload Protection: 200mV is 230V (AC or DC Current).
All other ranges are 500Vrms.
Display: Average Value (RMS of Sine Wave)
DC Current
Range Resolution Accuracy
2000µA 1µA
20mA 10µA
± (1.0% rdg + 2 digits)
200mA 100µA ± (1.2% rdg + 2 digits)
10A 10mA ± (2.0% rdg + 5 digits)
Overload Protection: 0.3A / 500V fuse, 10A range is unfused.
Max. Current Input: 10A (10 seconds max, 15 minute cool down)
Burden Voltage: 300mV max