Triplett 310 Manual de usuario Pagina 29

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 34
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 28
27
Test Leads:
79-153: 42” Test leads w/Screw-on Alligator Clips
79-206: 48” Test leads with Alligator Clips (no probes)
79-695: Test leads with Pin Cushion Clips
79-722: Test leads with Spike Clips
MAINTENANCE FOR
MODEL 310/310-C
Battery Replacement
If the pointer cannot be adjusted to full scale on the
X1, X10, or X100 OHMS ranges, replace the 1.5V bat-
tery. Replace the 12V battery if the X1K OHMS range
cannot be adjusted for full scale.
1.5V Battery - NEDA 910F or 910M (PN: 37-21)
12V Battery - A23 (PN: 37-60)
Fuse Replacement
If none of the OHMS ranges work, replace the fuse.
Use a 1/16 AMP, 5 mm x 20 mm fuse. (PN: 3207-112)
Cleaning Plastic Window
The plastic window has been treated at the factory to
dissipate static charges. If cleaning is necessary use
cotton dipped in a solution of common household
detergent and water. After cleaning, allow the solu-
tion to dry without rubbing; the resultant detergent
film will effectively dissipate static charges.
CAUTION:
Solven
ts and liquids used in radio and TV shop work
may craze or scar the plastic window if applied to it.
Vista de pagina 28
1 2 ... 24 25 26 27 28 29 30 31 32 33 34

Comentarios a estos manuales

Sin comentarios