7.8
Continuity
Beeper:
WARNING!
Do not apply voltage or current to the meter when it is set to
the Diode Test / Continuity Beeper range.
7.8.1 Connect the black test lead to the COM jack and the red
test lead to the input jack indicated as on the “Front
Panel” drawing on page 22.
7.8.2 Set the RANGE switch to the Diode Test / Continuity
Beeper“ range.
7.8.3 Connect the test probes to the device or circuit to be
tested.
7.8.4 If the resistance of the device or circuit is less than 20 to
100 Ohms, the continuity beeper will sound.
Note: The reading displayed on the LCD
is
not an accurate indication
of
the resistance
of
the device
or
circuit being
measured.