25
Frequency
Range Resolution Accuracy
5.000Hz 0.001Hz
50.00Hz 0.01Hz
500.0Hz 0.1 Hz
5.000kHz 1 Hz
50.00kHz 10Hz
500.0kHz 100Hz
± 1.2% of rdg ± 3 dgts
5.000MHz 1kHz
10.00MHz 10KHz
± 1.5% of rdg ± 4 dgts
Sensitivity: 0.5V RMS at frequencies less than 1MHz ; 3V
RMS at frequencies greater than 1MHz ;
Input Protection: 600V AC/DC
Duty Cycle
Range Resolution Accuracy
5 to 95% 0.1% +/- 5% rdg
Diode and Continuity
Range Test Current Other
Diode Test 0 to 1.000V
+/- 2% typ
Continuity
1.4mA, 2.8V
Open Circuit Voltage
Beeps if < 50
Overload Protection: maximum 600V dc or 600V ac rms.